Class 2 Device Recall Philips Healthcare Computed Tomography XRay System
Date Initiated by FirmApril 05, 2010
Date PostedApril 17, 2013
Recall Status1Open3, Classified
Z-1003-2013
Recall Event ID64708
System, x-ray, tomography, computed - Product CodeJAK
ProductPhilips Healthcare Computed Tomography X-Ray System.
These devices are whole-body computed tomography (CT) x-ray
systems or sub-systems, each with a continuously rotating x-ray tube
and multi-row detectors enclosed by a gantry. X-ray transmission
data acquired and taken at different angles can be reconstructed into
cross-sectional images. Each device also includes signal analysis
and display equipment, patient and equipment supports, components, and accessories.
Code InformationGemini Dual, Gemini 16 Power, Gemini GXL, Gemini LXL, Gemini TF 16, Gemini TF Base/Ready, Gemini TF 64, Gemini TF Big Bore, Precedence SPECT/CT, Brilliance CT (6, 10,16, 16P, 40, 64, and Big Bore), Brilliance iCT, Brilliance iCT SP
Recalling Firm/
Manufacturer
Philips Healthcare Inc.
3000 Minuteman Road
Andover MA 01810
For Additional Information Contact
978-687-1501
Philips Healthcare discovered the customer was performing CTDI measurements for facility accreditation and received results that were 15 to 25% higher than Philips protocol parameters stated.
Radiation Control for Health and Safety Act
Philips Healthcare proposed corrective action plan (CAP) is comprised of the following elements: * Updating procedures in manufacturing to verify, either through software calibration or through gage-pin measurement, that A-plane collimation is appropriate to the design specification. It is our understanding that these procedures have been implemented. * For refurbished PET/CT and SPECT/CT systems distributed by Philips, the manufacturing process now requires that CTDI testing be conducted in addition to air testing and that each system meet updated CTDI tolerance limits to be published in the product's instructions for use. * Reassessing and correcting the correlation between air dose and CTDI specifications for all CT systems or sub-systems, resulting in updates of the CTDI specifications (tolerances) to be published in the instructions for use. It is our understanding that this reassessment and correction have been implemented, that the updated CTDI specifications will be made available as new CTDI tolerance tables with the customer information letter to be distributed immediately, and that these updated CTDI specifications will also be distributed as an addendum to the instructions for use manuals to the installed base during the next 12 months. * For all Brilliance CT, Gemini PET/CT, and Precedence SPECT/CT systems, in the installed base as well as in forward production, instructions for field service will be updated to provide clear directions on how to verify that the A-plane collimator has been appropriately adjusted to ensure expected product performance after its replacement and/or the replacement of an x-ray tube. This plan appears to adequately address the problem, and it is hereby approved. For further questions please call (978) 687-1501.
2089 US 4036 ROW
Worldwide Distribution including USA (nationwide)
TPLC Device Report
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.

Philips Gemini Tf Operator Manual Transmission

Manuals and free owners instruction pdf guides. Find the user manual and the help you need for the products you own at ManualsOnline. More parts for Gemini TF 64, Gemini Big Bore, and Brilliance 64 coming soon; Call AllParts Medical at 866 507 4793 for access to the Full Philips Imaging Parts Catalog. As part of the Philips family, AllParts Medical has access to all the Philips MRI, CT, Ultrasound and Xray parts you need. GEMINI: GOODMANS: GRAETZ: GRANADA: GRUNDIG: HAMEG: HANSEATIC. Use search box below to find Your manual! Enter model or manufacturer. Philips - Gemini TF Astonish TF technology has enhanced PET imaging performance. Philips has now expanded imaging capabilities with its latest innovation &nda. Boost user procedure capacity and improve patient experience with fast PET acquisition times, as short as 10 minutes. Need service manual, Yellowing Epoxy. I need a service manual.